Clarke-Hess 8200 and 8210 Transconductance Amplifier
- Output current range: 200 µA - 100 A
- Frequency range:
8200: DC - 100 kHz
8210: DC - 10 kHz - Compliance voltage: 7 V RMS (AC) / 7 V (DC)
- Total harmonic distortion: -60 dB (up to 10 kHz)
- Short-term stability: 40 ppm
- Input impedance: 500 kΩ
- Inductive load stability: Free of oscillations up to 1 mH
- Interfaces: IEEE-488.2 (GPIB), USB
The Clarke-Hess 8200 and 8210 are high-precision transconductance amplifiers designed for applications in power electronics, test and measurement laboratories and calibration environments. These amplifiers convert an input voltage into a proportional output current with high accuracy and stability over a wide frequency range of DC to 100 kHz (8200) and DC to 10 kHz (8210). With 100A output current, a compliance voltage of 7 V RMS, and low harmonic distortion (-60 dB to 10 kHz), these amplifiers are ideal for testing current transformers, power sensors and power meters.
Excellence in performance and accuracy
The Clarke-Hess 8200 and 8210 are designed for applications requiring high precision, stability and minimal distortion:
- > Output current range: 200µA to 100A in six ranges
- > Frequency range:
- 8200: DC – 100 kHz
- 8210: DC – 10 kHz
- > Short-term stability: 40 ppm
- > DC accuracy: 0,04%
- > AC accuracy: 0,10%
- > Total harmonic distortion:
- < -60 dB (to 10 kHz)
- < -50 dB (to 50 kHz, 8200)
- < -40 dB (to 100 kHz, 8200)
The Clarke-Hess 8200 and 8210 maintain accuracy independent of load and can operate with inductive and capacitive loads without instability.
Versatile functionality
> 100% overload set range: All ranges can be loaded up to 200% of their rated value without loss of performance (except the 100 A range).
> Advanced colour display: Settings, compliance voltage (0 – 7 V) and input signal frequency (10 Hz – 500 kHz) are displayed in real time.
> Front panel calibration: DC offset and transconductance can be adjusted directly without opening the housing.
> Coaxial output for high currents: The 20 A and 100 A ranges use an LC connector to minimise inductance.
Applications
The Clarke-Hess 8200 and 8210 are ideal for:
> Calibration and testing of power meters, transducers and current transformers
> Research and development in power electronics
> Testing of precision resistors and low-resistance load tests
> Simulation of realistic current profiles for industrial and scientific applications
The Clarke-Hess 8200 and 8210 offer a unique combination of high precision, flexibility and ease of use, making them essential tools for current calibration and power electronics.